Model Number UNKNOWN
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Meningitis (2389)
Event Type  Injury  
Manufacturer Narrative
This investigation is currently ongoing.Any additional information will be provided in the follow-up report.
 
Event Description
According to the initial report, "the surgeon uses bioglue for expanded endonasal skull base procedures - pituitary, encephalocele repair, meningioma.Bioglue was used directly on dura after duraplasty in the posterior fossa; it caused a severe prolonged chemical meningitis.
 
Manufacturer Narrative
Additional information received indicates that the patient recovered from the meningitis; no long term consequences of the bioglue use; surgery in 2015.A re-do surgery was performed to remove bioglue and supportive care provided.It was also reported that the surgeon has used and will continue to use bioglue for her transphenoidal procedures, the patient in question made a good recovery with no ill effects from the use of the glue.This report is being submitted as required by federal regulations and does not constitute an admission that the device caused or contributed to the reported event.Furthermore, this report reflects the event as alleged by the complainant and does not imply that the information reported to cryolife is accurate or has been confirmed by cryolife.
 
Event Description
According to the initial report, "the surgeon uses bioglue for expanded endonasal skull base procedures - pituitary, encephalocele repair, meningioma.Bioglue was used directly on dura after duraplasty in the posterior fossa, it caused a severe prolonged chemical meningitis.
 
Event Description
According to the initial report, "the surgeon uses bioglue for expanded endonasal skull base procedures - pituitary, encephalocele repair, meningioma.Bioglue was used directly on dura after duraplasty in the posterior fossa, it caused a severe prolonged chemical meningitis.
 
Manufacturer Narrative
Based on the information available at the time of this report, we are unable to definitively determine the root cause of the chemical meningitis observed in this patient or if bioglue, the patch material, and/or suture contributed to the observed event.It is possible that the surgeon used the bioglue alone, rather than as an adjunct to sutures, which could result in application of bioglue into unintended areas such as intradurally.Intradural bioglue could potentially cause an inflammatory reaction exhibited as chemical meningitis.There is no indication that an error or deficiency occurred at cryolife.The risk management file thoroughly identifies the process and product hazards for approved indications.Each individual hazard is mitigated and reduced as low as possible by design and process.The ifu (instructions for use) indicates the following potential adverse events may occur with the use of bioglue: inflammatory, immune systemic allergic reaction.This event does not identify additional hazards or modify the probability and severity of existing hazards.This report is being submitted as required by federal regulations and does not constitute an admission that the device caused or contributed to the reported event.Furthermore, this report reflects the event as alleged by the complainant and does not imply that the information reported to cryolife is accurate or has been confirmed by cryolife.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOGLUE SURGICAL ADHESIVE
Type of Device
GLUE, SURGICAL, ARTERIES
Manufacturer (Section D)
CRYOLIFE, INC.
1655 roberts blvd., nw
kennesaw GA 30144
MDR Report Key8212213
MDR Text Key131912720
Report Number1063481-2019-00001
Device Sequence Number1
Product Code MUQ
Combination Product (y/n)N
PMA/PMN Number
P010003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 03/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Model NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Distributor Facility Aware Date12/05/2018
Initial Date Manufacturer Received 12/05/2018
Initial Date FDA Received01/03/2019
Supplement Dates Manufacturer Received12/05/2018
12/05/2018
Supplement Dates FDA Received02/19/2019
03/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other;
-
-