• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRISMATIK DENTALCRAFT, INC. HAHN TAPERED IMPLANT DRIVER Ø5.0 SHORT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRISMATIK DENTALCRAFT, INC. HAHN TAPERED IMPLANT DRIVER Ø5.0 SHORT Back to Search Results
Model Number 70-1071-SRG0086
Device Problem Accessory Incompatible (1004)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/01/2018
Event Type  malfunction  
Manufacturer Narrative
Date of event was asked; however, the doctor advised it was some time in (b)(6) 2018.Patient's date of birth and age were asked; however, the information has not been provided.Follow-ups were made to request more information; however, some information are still pending from the doctor.The reported device was received and the evaluation is anticipated.Once the investigation is completed, a supplemental report will be submitted.This report for is the 3rd of 3 reported implant drivers: please reference 3011649314-2018-00453 ((b)(6)) for the 1st implant driver.Please reference 3011649314-2018-00454 ((b)(6)) for the 2nd implant driver.
 
Event Description
This report is for the 3rd of 3 failed implant drivers: it was reported that a short hahn tapered implant driver ø5.0 failed to engage properly.Per provided information, the implant driver would spin freely even though the doctor reported proper size implant driver was used.The doctor could not reach the desired depth or could not back the implant out to re-drill.There was no known injury to the patient.
 
Manufacturer Narrative
The reported driver was returned for evaluation.A visual and microscropic inspection were performed on the returned driver.The critical parameters of the driver were measured and no defects, nor non-conformities were found.A lot number was received and a device history record review (dhr) was conducted.There was no evidence to indicate that a product defect or nonconformity contributed to the issue.The driver met all the criteria called for in the production router.A fit test was performed on the driver using a stocked implant.The driver could be fully seated inside the implant.There was no engagement issue observed between the driver and stocked implant.A second fit test was also performed on the driver using both a stocked handpiece and a stocked surgical wrench.The driver latched onto the handpiece and surgical wrench without any issue.There was no engagement issue found with the driver to the handpiece nor the surgical wrench.The driver did not "spin freely" during testing.Based on the investigation, the root cause could not be determined as the driver functioned as intended.Without the associated implant returned, the exact replication of the event could not be performed.The reported issue could not be confirmed.This event will be tracked and trended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAHN TAPERED IMPLANT DRIVER Ø5.0 SHORT
Type of Device
HAHN TAPERED IMPLANT DRIVER Ø5.0 SHORT
Manufacturer (Section D)
PRISMATIK DENTALCRAFT, INC.
2212 dupont drive
suite p
irvine CA 92612
MDR Report Key8215658
MDR Text Key132083898
Report Number3011649314-2018-00455
Device Sequence Number1
Product Code NDP
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number70-1071-SRG0086
Device Catalogue Number70-1071-SRG0086
Device Lot Number6026304
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2018
Initial Date Manufacturer Received 12/04/2018
Initial Date FDA Received01/03/2019
Supplement Dates Manufacturer Received04/02/2019
Supplement Dates FDA Received04/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Weight73
-
-