MAUDE Adverse Event Report: BAXTER HEALTHCARE ¿ ROUND LAKE 250 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE ¿ ROUND LAKE 250 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938737
Device Problems Fluid/Blood Leak (1250); Material Perforation (2205)
Patient Problem No Patient Involvement (2645)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 250ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag had a pin-sized hole in the middle of the bag resulting in leakage.The leak was discovered prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information : the lot number was manufactured between november 03, 2017 - november 04, 2017.The actual device was not available; however, a photograph of the sample was provided for evaluation.The returned photograph showed evidence of a leak coming from the administration port connector.The reported condition was verified.The cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.This issue is being further investigated.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
250 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE ¿ ROUND LAKE
round lake IL
Manufacturer (Section G)
BAXTER HEALTHCARE ¿ ROUND LAKE
25212 w. illinois route 120
round lake IL 60073
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8219643
MDR Text Key132199458
Report Number1416980-2018-08199
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477213
UDI-Public(01)00085412477213
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date10/31/2020
Device Catalogue NumberH938737
Device Lot Number60095447
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/12/2018
Initial Date FDA Received01/04/2019
Supplement Dates Manufacturer Received01/25/2019
Supplement Dates FDA Received02/07/2019
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-