ZOLL CIRCULATION INC AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 12/18/2018
Event Type  malfunction  
Manufacturer Narrative
Zoll has not received the autopulse platform (sn (b)(4)) for investigation.A supplemental report will be filed when the product is returned and the investigation has ben completed.Please see the following related mfr report: mfr # 3010617000-2019-00016 for the lifeband.
 
Event Description
As reported, during shift check, the lifeband was unable to be installed to the autopulse platform (sn (b)(4)).No patient involvement.
 
Manufacturer Narrative
The reported event was confirmed through visual inspection of the autopulse platform (sn # (b)(4).The root cause is due to a sticky encoder drive shaft.Upon visual inspection, no physical damages were observed.Noticed that the encoder drive shaft does not rotate smoothly, exhibits binding and resistance.The clutch plate was deburred to remedy the fault.During initial functional testing, the platform passed the testing without any fault / error.The archive data review showed no errors.Following the service, the autopulse was subjected to the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test batteries until discharged without any fault or error.The brake gap inspection was performed and verified the brake gap was within the specification.Load cell characterization test was performed and confirmed both cell modules are functioning within the specification.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse platform with serial number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION INC
2000 ringwood avenue
san jose CA 95131
MDR Report Key8232776
MDR Text Key132576120
Report Number3010617000-2019-00015
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111065009
UDI-Public00849111065009
Combination Product (y/n)N
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 01/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-03
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/24/2018
Initial Date Manufacturer Received 12/20/2018
Initial Date FDA Received01/10/2019
Supplement Dates Manufacturer Received01/15/2019
Supplement Dates FDA Received01/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-