Model Number 24658 |
Device Problems
Leak/Splash (1354); Material Rupture (1546)
|
Patient Problem
Perforation of Vessels (2135)
|
Event Date 12/18/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.
|
|
Event Description
|
It was reported that patient experienced minor vascular perforation.The 80% stenosed target lesion was located at the vein shunt.A 4.00mm/1.5cm/140cm small peripheral cutting balloon was selected fro use.During the procedure, at second dilatation, it was noted that balloon pinhole rupture occurred at 12atm and the patient experienced minor perforation from the blood vessel.The procedure was completed after hemostasis with a non bsc device.No further patient complications were reported and the patient was stable after treatment.
|
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.Device evaluated by mfr: product analysis identified a balloon pinhole and hypotube kinks.This type of damage is consistent with excessive force being applied to the device.There is no evidence that the device failed to meet spcb flextome specification prior to shipping.The use of this flextome device indicates the requirement for dilatation and lesion scoring.The device encountered difficult patient anatomy as it was stated that there were two lesions with 80% stenosis, which would have contributed to the complaint incident.As there were no kinks reported in the event, the kinks on the hypotube may have occurred when the device was being returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was identified within the balloon and inflation lumen which is evidence of a device leak.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 9mm distal to the distal end of the proximal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination observed no damage to the tip or blades.All blades were present and fully bonded to the balloon surface.
|
|
Event Description
|
It was reported that patient experienced minor vascular perforation.The 80% stenosed target lesion was located at the vein shunt.A 4.00mm/1.5cm/140cm small peripheral cutting balloon was selected fro use.During the procedure, at second dilatation, it was noted that balloon pinhole rupture occurred at 12atm and the patient experienced minor perforation from the blood vessel.The procedure was completed after hemostasis with a non bsc device.No further patient complications were reported and the patient was stable after treatment.
|
|
Search Alerts/Recalls
|