• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED URO-SOLUTIONS, L.L.C. STIMLOC; COVER, BURR HOLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED URO-SOLUTIONS, L.L.C. STIMLOC; COVER, BURR HOLE Back to Search Results
Model Number 924256
Device Problem Unstable (1667)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/09/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider (hcp) regarding a patient who was implanted with a neurostimulator (ins) for movement disorders.It was reported that the inner clip of the stimloc was not holding the dbs lead.After clipping the lead it pulled back a couple of cm and had to be put back in place.O arm scan at the end confirmed that the lead was relocated to the correct depth.The clip was repositioned but was still not holding the lead as well as usual.The stimloc cal was then put in to further secure the lead.The issue was resolved.There were no further complications reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from the manufacturer¿s representative (rep), which was confirmed with the healthcare provider (hcp), reported there was no reason for the clip not holding.The clip was still implanted and would not be returned.No further patient symptoms or complications were anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STIMLOC
Type of Device
COVER, BURR HOLE
Manufacturer (Section D)
ADVANCED URO-SOLUTIONS, L.L.C.
7842 hickory flat highway
suite d
woodstock GA 30188
Manufacturer (Section G)
ADVANCED URO-SOLUTIONS, L.L.C.
7842 hickory flat highway
suite d
woodstock GA 30188
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8237428
MDR Text Key132941172
Report Number3012165443-2019-00001
Device Sequence Number1
Product Code GXR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/11/2021
Device Model Number924256
Device Catalogue Number924256
Device Lot Number082225418A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/09/2019
Initial Date FDA Received01/10/2019
Supplement Dates Manufacturer Received01/10/2019
Supplement Dates FDA Received01/17/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient Weight60
-
-