MAUDE Adverse Event Report: MEDTRONIC NAVIGATION ISRAEL, LTD POLESTAR N-30 MRI SYSTEM; SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION ISRAEL, LTD POLESTAR N-30 MRI SYSTEM; SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING Back to Search Results
Model Number 9734070
Device Problems Computer Software Problem (1112); Patient Device Interaction Problem (4001)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/20/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was not known at the time of filing.Device udi not known at the time of filing.No devices were returned to the manufacturer for analysis.Device manufacturing date is not known at the time of filing.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging device being used pre-operatively of a cranial resection procedure.It was reported that the t2 flair scan was interrupted and aborted.The system did not work properly.It was later noted that the issue was resolved after re-booting the system 3 times.The total delay of the procedure was longer than 1-hour and no impact on patient outcome.
 
Manufacturer Narrative
A medtronic representative went to the site to inspect the system.The system passed all test and performed as intended.No parts required replacement and no changes were made to the system configuration files.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLESTAR N-30 MRI SYSTEM
Type of Device
SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING
Manufacturer (Section D)
MEDTRONIC NAVIGATION ISRAEL, LTD
3 hacarmel street p.o. box 548
kochav yokneam building
yokneam elit ha zafon
MDR Report Key8246148
MDR Text Key133060732
Report Number9710108-2019-00001
Device Sequence Number1
Product Code LNH
Combination Product (y/n)N
PMA/PMN Number
K092308
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9734070
Device Catalogue Number9734070
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/20/2018
Initial Date FDA Received01/14/2019
Supplement Dates Manufacturer Received01/03/2019
Supplement Dates FDA Received02/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-