Catalog Number 320-15-01
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Joint Disorder (2373)
Event Type  Injury  
Manufacturer Narrative
In a review of the labeling it is a known complication that a patient's age, weight, or activity level would cause the surgeon to expect early failure of the system.Revisions or surgical interventions are a known complication found in joint replacements.Based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of instability of the left shoulder joint devices is most likely related to the patient's underlying conditions to include previous significant shoulder maladies.This device is used for treatment, not diagnosis section(s): no information: serial and expiration date.
 
Event Description
It was reported through clinical data collection activities that a patient experienced device instability to the left shoulder with an onset of (b)(6) 2018 that required revision surgery on (b)(6) 2018.The revision marked the event as being as "resolved".The patient had described a history of at least 3-5 anterior instability, as subluxations that were self-reduced.The patient has significant medical history for left shoulder surgeries and/or injuries.There is no indication that the device malfunctioned.No additional information is provided.This is one of five products involved with the reported event and the associated manufacturer report numbers are 1038671-2018-00471, 1038671-2019-05000, 1038671-2019-05001, 1038671-2019-05002 and 1038671-2019-05003.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE REVERSE GLENOID PLATE
Type of Device
REVERSE GLENOID
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
graham cuthbert
2320 nw 66th court
gainesville, FL 32653
MDR Report Key8249046
MDR Text Key133180845
Report Number1038671-2019-05003
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063569
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type study
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Catalogue Number320-15-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/14/2019
Initial Date FDA Received01/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age56 YR
Patient Weight95
-
-