ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER LAUREATE WORLD PHACO SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750502
Device Problem Intermittent Continuity (1121)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that after a procedure, the system was shutting down on its own intermittently.
 
Manufacturer Narrative
The company service representative examined the system and was able to confirm and replicate the reported event.The nonconforming power supply was replaced.The system was then tested and met all product specifications.The system was manufactured on december 18, 2014.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to the nonconforming power supply.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAUREATE WORLD PHACO SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key8250351
MDR Text Key133209882
Report Number2028159-2019-00093
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K021566
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750502
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/18/2018
Initial Date FDA Received01/15/2019
Supplement Dates Manufacturer Received03/13/2019
Supplement Dates FDA Received03/28/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-