• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 SELF CENT HIP 46X28 BRN; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 SELF CENT HIP 46X28 BRN; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI Back to Search Results
Catalog Number 103546000
Device Problem Fitting Problem (2183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/19/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).No device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the polyethelyne was loose in the bipolar head upon unboxing the implant.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.: product complaint # (b)(4).Investigation summary: updated: (b)(4), 2019.The device has been returned for examination.Examination of the returned device components is unable to confirm the reported problem.The components can be assembled as intended with no excess movement.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF CENT HIP 46X28 BRN
Type of Device
HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8252217
MDR Text Key133196554
Report Number1818910-2019-81036
Device Sequence Number1
Product Code KWY
UDI-Device Identifier10603295003366
UDI-Public10603295003366
Combination Product (y/n)N
PMA/PMN Number
K812672
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number103546000
Device Lot NumberJ15W29
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2019
Initial Date Manufacturer Received 12/19/2018
Initial Date FDA Received01/16/2019
Supplement Dates Manufacturer Received02/25/2019
04/15/2019
Supplement Dates FDA Received02/26/2019
04/15/2019
Patient Sequence Number1
Patient Age90 YR
-
-