If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).The complaint device is not being returned, therefore unavailable for a physical evaluation.A review into the depuy mitek complaints system revealed no other complaints for this lot of devices that were released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.This file will remain receptive to any potential forthcoming information received that is pertinent and germane to this issue.A non-conformance search was performed for this product code 231814, lot #3897528 combination and no non-conformances were identified.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the etq complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
It was reported by the affiliate in (b)(6) that during an unspecified surgical procedure, it was observed that the 5x12 milagro interference screw device fell out of the bone after it has been placed.According to the reporter, the insertion was not off axis.It was reported that the bone quality was hard.It was reported that the procedure was completed using the same bone hole there was a delay of 15 minutes.The procedure was completed with same like product.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|