Results: the jet7 was partially fractured approximately 47.0 cm from the hub.The device was ovalized approximately 56.5 cm from the hub.The jet7 was flushed with air inside of the bleach solution and air bubbles were seen exiting the device at the fractured location.Conclusions: evaluation of the returned jet7 confirmed a leak on the midshaft.The returned device was kinked at the location of the leak.The reported multiple passes with the device may have contributed to the damaged device.This damage typically occurs due to forcefully advancing the device against resistance or manipulating at extreme angles outside of the patient.The root cause of the kink could not be determined.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure in the middle cerebral artery (mca) using a penumbra system jet7 kit.During the procedure, after making two passes in the target vessel using a penumbra system jet 7 reperfusion catheter (jet7) with a neuron max 6f 088 long sheath (neuron max), the hospital technician noticed a leak in the mid shaft while flushing the jet7 with saline.Therefore, it was removed.The procedure was completed using a new jet7 with the same sheath.There was no report of an adverse effect to the patient.
|