• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN SHARPSAFETY; CONTAINER, SHARPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN SHARPSAFETY; CONTAINER, SHARPS Back to Search Results
Model Number 8980S
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/16/2018
Event Type  malfunction  
Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer reported the sliding lid can be reversed but will not lock.
 
Manufacturer Narrative
Submission date: 02/11/2019.No lot number was provided.Therefore, a review of the device history record was unable to be performed.However, all device history records (dhr) are reviewed for accuracy prior to product release.In-process procedures are also in place to prevent nonconforming product in the manufacturing process.This ensures components and finished products meet all quality inspection standards.These controls include, but are not limited to: material verification/certification processes, dimensional specifications, statistical samplings, periodic audits, process inspections, machine maintenance/operation and personnel training and certification.One (1) sample was received for evaluation.Sample received had 8g sliding top (clear) reversed.Side panel is cracked, where customer attempted to correct.The reported customer complaint is confirmed.The root cause was determined to be the operator miss-assembled sliding lid in reverse position.This complaint will be used for tracking and trending purposes.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHARPSAFETY
Type of Device
CONTAINER, SHARPS
Manufacturer (Section D)
COVIDIEN
815 tek dr
crystal lake IL 60014 8172
MDR Report Key8264259
MDR Text Key133715396
Report Number1424643-2019-00001
Device Sequence Number1
Product Code MMK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8980S
Device Catalogue Number8980S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2019
Initial Date Manufacturer Received 12/24/2018
Initial Date FDA Received01/21/2019
Supplement Dates Manufacturer Received02/07/2019
Supplement Dates FDA Received02/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-