• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS VOLISTA; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS VOLISTA; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARD568811960
Device Problem Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/28/2018
Event Type  malfunction  
Manufacturer Narrative
The issue is being investigated by manufacturer.(b)(4).
 
Event Description
On (b)(4) 2018 maquet (b)(4) became aware about an issue with one of the surgical lights- volista.As it was stated, the paint on fork of the device was damaged.There was no injury reported and related to this issue.However we decided to report the issue in abundance of caution as such damage in the paint integrity can result in parts falling into the sterile field and might be a source of contamination.(b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.Getinge usa sales, llc (importer) is submitting this report on behalf of the legal manufacturer of the device maquet sas, parc de limère, avenue de la pomme de pi orléans cedex 2, france 45074 exemption # e2018005.Getinge usa sales, llc 45 barbour pond drive wayne, nj 07470.Contact person: (b)(4).
 
Manufacturer Narrative
Getinge usa sales, llc (importer) is submitting this report on behalf of the legal manufacturer of the device maquet sas, parc de limère, avenue de la pomme de pi orléans cedex 2, france 45074 exemption # e2018005.Getinge usa sales, llc, 45 barbour pond drive, wayne, nj 07470.Contact person: (b)(6).Maquet sas became aware of a customer¿s problem with the volista device.As it was stated by the technician, there were cracks in the paint observed, specifically at the arc of the fork.There was no injury to a patient reported, however it was decided to report this issue based on the potential and in abundance of caution, as any particle falling from the device into the sterile field might be a source of contamination.When reviewing similar reportable events registered for volista surgical light we were able to find several similar issues compared to the problem investigated herein.In none of the complaints a serious injury or death occurred.Based on the information collected to date it was established that when the event occurred, the surgical light did not meet the manufacturer¿s specification.The most probable root cause that the cracks appear was the detachment of the bonded parts of fork.It was established that the cracks were located only on the outer coating and there was no impact to the internal structure of the fork assembly.The breaks on the coating correspond to the excessive gap between two mechanical parts under the outer material.We have no information regarding the exact time when the defect first appeared or if it was being used for patient treatment in the time when the event occurred.A corrective/preventive action investigation into the issue has revealed that the issue may be the result of several factors, summarized as issues with the previous understanding of post-gluing aspects.A design change improved the assembling methods from glue-bonding to welding of the parts in production, since beginning of 2017.
 
Event Description
Manufacturer reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLISTA
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
orléans cedex 2
FR 
MDR Report Key8267123
MDR Text Key133844753
Report Number9710055-2019-00017
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
PMA/PMN Number
K130513
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Inspection
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 04/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberARD568811960
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/28/2018
Initial Date FDA Received01/22/2019
Supplement Dates Manufacturer Received12/28/2018
12/28/2018
12/28/2018
12/28/2018
Supplement Dates FDA Received02/12/2019
03/08/2019
04/02/2019
04/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-