Internal file number - (b)(4).The incident information was reviewed; however, the product was not returned to abbott vascular for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported/noted difficulty to position, to remove appear to be related to circumstances of the procedure.It is likely inadvertent mishandling of the device during preparation or during loading of the wire the distal shaft of the device became kinked resulting in the reported difficulty to position the wire.Additional manipulation of the wire and device likely caused the wire to separate causing the reported difficulty to remove.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
|