• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MANUFACTURING INC CLINICAL CHEMISTRY MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MANUFACTURING INC CLINICAL CHEMISTRY MAGNESIUM Back to Search Results
Catalog Number 07D70-21
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
A product correction letter was issued to all magnesium customers who have received one of the impacted lots.The letter instructs the customer to discontinue use of the magnesium urine application.New customers will receive product with a kit stuffer that includes instructions to immediately discontinue use of the 7d70 magnesium urine application.
 
Event Description
The customer reported magnesium urine controls shifted low out of range while using the architect urine magnesium assay.During troubleshooting the customer retested previously tested cap survey samples and all the mg urine results were decreased from the original results.No impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINICAL CHEMISTRY MAGNESIUM
Type of Device
MAGNESIUM
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer (Section G)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key8268894
MDR Text Key134717345
Report Number1628664-2019-00045
Device Sequence Number1
Product Code JGJ
UDI-Device Identifier00380740012922
UDI-Public00380740012922
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K981791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial
Report Date 01/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/07/2020
Device Catalogue Number07D70-21
Device Lot Number95220UN18
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/11/2019
Initial Date FDA Received01/22/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/13/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction Number1628664-01/8/19-001-C
Patient Sequence Number1
-
-