BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL SALINE SYRINGE; FLUSH SYRINGE Back to Search Results
Catalog Number 306595
Device Problem Difficult to Advance (2920)
Patient Problem No Information (3190)
Event Date 01/07/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation: four samples were received for evaluation.Breakout force and sustaining force testing results showed they all are within specification.Based on these results the complaint can't be confirmed and a root cause is undetermined.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.
 
Event Description
It was reported with the use of the bd posiflush¿ normal saline syringe there were 6 occurrences of when the plunger was pushed halfway it could not be pushed forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NORMAL SALINE SYRINGE
Type of Device
FLUSH SYRINGE
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key8270576
MDR Text Key133937561
Report Number1911916-2019-00105
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K982558
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date02/28/2021
Device Catalogue Number306595
Device Lot Number8046825
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/07/2019
Initial Date FDA Received01/23/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/15/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-