• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 806455
Device Problems Component Missing (2306); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/16/2018
Event Type  malfunction  
Manufacturer Narrative
Evaluation in progress but not yet concluded.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, while calibrating the occluder head, there was a noise coming from it and there were also screws missing.The electronic occluder was replaced for a manual occluder.The surgical procedure was completed successfully.There was a delay of five minutes.There was no blood loss nor adverse consequences to the patient.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) observed a louder than normal noise coming from the motor of the occluder head, especially at full occlusion.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
Per clinical review: during set up for a cardiopulmonary bypass (cpb) procedure on (b)(6) 2018, the clinicians had an incident with the venous occluder.When the clinician was calibrating the occluder there was a high pitched noise coming from inside the occluder, therefore they deemed the occluder head to be malfunctioning and opted to exchange the occluder out.The delay in set up was approximately five minutes.There was no blood loss or harm associated with the delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key8271652
MDR Text Key133890749
Report Number1828100-2019-00037
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 03/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number806455
Device Catalogue Number806455
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/03/2019
Initial Date FDA Received01/23/2019
Supplement Dates Manufacturer Received02/07/2019
03/06/2019
Supplement Dates FDA Received03/04/2019
03/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-