The patient was in ipp 9 for a coiling procedure.During the case, the md noticed changes in the image quality in the regards to the roadmap shifting even without the breath of the patient.Routes were taken to try and mitigate the issue, but they were unsuccessful in fixing the issue.To avoid wasting time or possibly rupture the aneurysm due to poor imaging, the choice was made to move the whole case to ipp 7.It was also a concern since more scenes, roadmaps, and imaging was needed to clarify the location of the micro catheter, resulting in a high patient dose.
|