MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS VITAMIN D TOTAL II; VITAMIN D TEST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS VITAMIN D TOTAL II; VITAMIN D TEST SYSTEM Back to Search Results
Catalog Number 07464215190
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/03/2019
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer received questionable vitamin d total g2 results from cobas e411 disk analyzer serial number (b)(4).The initial result was 73.59 ng/ml and repeat results were 13.81 ng/ml and 12.90 ng/ml.No erroneous result was reported outside of the laboratory.There was no allegation of an adverse event.
 
Manufacturer Narrative
After further investigation it was determined that serum samples are not affected by the recall.Correction/removal report number has been updated.
 
Manufacturer Narrative
Roche diagnostics has issued an urgent medical device correction for this issue, entitled "elecsys® vitamin d total ii assay non-reproducible false high results." this correction has been reported to fda.During the implementation of the elecsys vitamin d total ii assay on the modular analytics e 170 module, and cobas e 601 and 602 systems, there were reports of non-reproducible, false high results.These customer observations were made in comparisons of duplicate measurements during assay validation of elecsys vitamin d total ii assay or in method comparisons with elecsys vitamin d assay (i.E., during conversions), where the falsely elevated discrepant value did not fit the expected result.The observed elevated results reported with the elecsys vitamin d ii assay were not confirmed upon repeat testing of the affected samples.The observed findings: the first result is elevated, either above the upper end of the measuring range (>100 ng/ml or >250 nmol/ml), or within the measuring range; repeated results are significantly lower.Rare cases have been reported on the cobas e 411 analyzer and the cobas e 801 module.Roche is conducting investigations into the reported issue and has determined that the elecsys® vitamin d total ii assay is strongly affected by pre-analytical errors.The investigations have reproduced these findings when requirements for pre-analytical sample handling have not been met for plasma samples.Further investigations into the reported issue are ongoing.As a result, the pre-analytical sample quality and compliance to the tube manufacturer¿s specifications is very important to assure a high quality sample in order to minimize the risk of occurrence.A workaround has been provided to customers.The calibration and qc data did not indicate an issue.The alarm trace did not indicate an issue.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS VITAMIN D TOTAL II
Type of Device
VITAMIN D TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key8274711
MDR Text Key134082949
Report Number1823260-2019-00248
Device Sequence Number1
Product Code MRG
Combination Product (y/n)N
PMA/PMN Number
K113546
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup,Followup
Report Date 07/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Catalogue Number07464215190
Device Lot Number32641201
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/06/2019
Initial Date FDA Received01/24/2019
Supplement Dates Manufacturer Received01/06/2019
01/06/2019
Supplement Dates FDA Received02/14/2019
07/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberRES 8008
Patient Sequence Number1
-
-