• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problems Intermittent Continuity (1121); Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/10/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported the system shutdown during a retinal surgical procedure following 5 system messages.The system was rebooted to complete the procedure with no harm to the patient.
 
Manufacturer Narrative
The customer called the company representative to assist with troubleshooting.The footswitch was determined to have been the cause the issue.The customer was able to resolve the issue remotely, by ordering a new footswitch.The footswitch was manufactured on april 30, 2012.Based on qa assessment, the product met specifications at the time of release.The root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key8276079
MDR Text Key134072618
Report Number2028159-2019-00136
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/10/2019
Initial Date FDA Received01/24/2019
Supplement Dates Manufacturer Received04/03/2019
Supplement Dates FDA Received04/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-