The reference device was not returned to olympus for evaluation.The exact cause of the reported event cannot be determined.However, the instruction manual warns users ¿introduce a new plasmakinetic supersect/loop device and the instrument cable into the sterile field.Insert the t-shaped connector on the instrument cable into the slot on the base of the white instrument mounting block.Ensure that the connector is fully seated within the block.¿ in addition , the oem has attributed this type of phenomenon to fluids invading the connection block causing an electrical short to occur due to the poor connection of the electrode, working element and cable.
|
This supplemental report is being submitted to provide the device evaluation results.A visual inspection of the received condition of the device found charring deposits inside electrode openings of the actuation block and cracks on o-ring inside the rack support.The device was then checked with olympus test electrode, active cord, and generator.There was no no spark or smoke observed during activation.In addition, the device passed the leak test.The cause of the reported event could not be conclusively determined as evaluation could not confirm the reported event and the concomitant devices that were used in the procedure were not returned along with the device for evaluation.Based on the similar reported complaint, the instruction manual provides instruction and warning that the device is to be used with a standard monopolar generator usa elite system, and recommended electrodes by oem.Also, if incomplete/improper assembly of the electrode to the active cord connector, are not locked in place, and resulted in an intermittent connection that could have produced arcing or sparking from the device.
|