• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL SALINE SYRINGE; VASCULAR CATHETER FLUSH SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL SALINE SYRINGE; VASCULAR CATHETER FLUSH SYRINGE Back to Search Results
Catalog Number 306546
Device Problems Crack (1135); Melted (1385); Material Twisted/Bent (2981); Sharp Edges (4013)
Patient Problem No Information (3190)
Event Date 01/16/2019
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that a bd posiflush normal saline syringe was bent, melted, sharp, and partially cracked upon opening the package.
 
Manufacturer Narrative
Investigation summary: one sample was received for evaluation.This is the first complaint for the lot# 8242677 for the same defect or symptom.There was no documentation of issues for the complaint of batch #8242677 during this production run.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.Investigation conclusion: the luer tip is damaged/ deformed therefore failure mode is verified.No other damages were observed on the barrel, plunger rod or tip cap.Root cause description: if the tip cap cup somehow has a piece of plastic from a tip cap jam, then that piece will induce the damage to the luer.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that a bd posiflush¿ normal saline syringe was bent, melted, sharp, and partially cracked upon opening the package.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NORMAL SALINE SYRINGE
Type of Device
VASCULAR CATHETER FLUSH SYRINGE
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
MDR Report Key8283799
MDR Text Key134363060
Report Number1911916-2019-00126
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public30382903065463
Combination Product (y/n)N
PMA/PMN Number
K141311
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/31/2021
Device Catalogue Number306546
Device Lot Number8242677
Initial Date Manufacturer Received 01/22/2019
Initial Date FDA Received01/28/2019
Supplement Dates Manufacturer Received01/22/2019
Supplement Dates FDA Received02/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-