TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS 8000 UNIVERSAL ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER
|
Back to Search Results |
|
Model Number 16402 |
Device Problem
Device Slipped (1584)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/07/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the srt, the belt was found to be off of the pulleys due to no spring tension from the motor tensioner.The belt was replaced.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
The service repair technician (srt) reported that during routine testing of the device at the service center, the pump produced a belt slip error.There was no patient involvement.
|
|