The patient was undergoing a thrombectomy procedure in the femoral vein using an indigo system separator 8 (sep8).During the procedure, the physician used an indigo system aspiration catheter 8 (cat8) with the sep8 for about 20 minutes.The sep8 was then removed to clean out the cat8 because the aspiration had slowed down.As the physician removed the sep8 from the cat8, it was noticed that the tip of the sep8, distal of the bulb, was unraveled.Therefore, the procedure was completed using a new sep8 and the same cat8.There was no report of an adverse effect to the patient.
|
Results: the indigo system separator 8 (sep8) core wire was fractured distal to the bulb approximately 149.5 cm from the proximal end.The wrapping wire winds distal to the bulb were stretched.The total length of the device was approximately 154.0 cm.Conclusions: evaluation of the returned sep8 confirmed the wrapping wire distal to the bulb was stretched and the core wire was fractured.If the device is manipulated repeatedly through tough clot burden, damage such as a fractured core wire may occur.If the core wire is fractured and the device is subsequently manipulated through clot burden or retracted from the patient, the wrapping wire will likely stretch.Penumbra separators are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|