A thorough evaluation of the x7-2t model transducer identified damage to the device.Visual inspection of this probe revealed several pin holes in the sheath, protein build up, chipped beading, a scratched i-tube, discolored cable jacket and dried fluid ingress in the connector.Further performance testing also found the steering cable ferrule had insufficient solder resulting in the failure to articulate in the anterior direction.The extensive physical damage to the transducer inhibited the overall performance of the device and is indicative of improper maintenance.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|