• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ABSORBATACK; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ABSORBATACK; STAPLE, IMPLANTABLE Back to Search Results
Model Number ABSTACK15
Device Problems Failure to Advance (2524); Detachment of Device or Device Component (2907)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/15/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter during a transabdominal preperitoneal procedure, the tacks of the first two firings came out at an angle.After that, some tacks came out as usual but also some tacks did not, then a new product was taken out and used.The part fell into the patient's cavity and was retrieved.There was no patient harm.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection and functional testing confirmed there were no abnormalities that would have caused or contributed to the reported condition.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSORBATACK
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key8303551
MDR Text Key134974464
Report Number1219930-2019-00697
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884523006506
UDI-Public10884523006506
Combination Product (y/n)N
PMA/PMN Number
K071920
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 03/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model NumberABSTACK15
Device Catalogue NumberABSTACK15
Device Lot NumberN8B0063MX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2019
Initial Date Manufacturer Received 01/15/2019
Initial Date FDA Received02/04/2019
Supplement Dates Manufacturer Received02/25/2019
Supplement Dates FDA Received03/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-