• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER RAPIDVAC; APPARATUS, EXHAUST, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER RAPIDVAC; APPARATUS, EXHAUST, SURGICAL Back to Search Results
Model Number SE3695
Device Problem Device Remains Activated (1525)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, prior to use, the device went into a high vacuum mode as soon as it was turned on and could not be stopped.There was no patient involved.
 
Manufacturer Narrative
Evaluation summary: one device was received for evaluation.A service request and a visual inspection and functional test were performed by the field service team.Inspection of the device resulted in verifying the issue.The device failed the performance test.This complaint will be considered as confirmed.A root cause analysis could not identify a root cause.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDVAC
Type of Device
APPARATUS, EXHAUST, SURGICAL
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow drive
boulder CO 80301 3299
MDR Report Key8305851
MDR Text Key135035737
Report Number1717344-2019-00157
Device Sequence Number1
Product Code FYD
UDI-Device Identifier10884524001753
UDI-Public10884524001753
Combination Product (y/n)N
PMA/PMN Number
K142335
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSE3695
Device Catalogue NumberSE3695
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2019
Initial Date Manufacturer Received 01/11/2019
Initial Date FDA Received02/05/2019
Supplement Dates Manufacturer Received02/20/2019
Supplement Dates FDA Received03/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-