DEPUY MITEK LLC US GRYPHON P BR DS ANCHOR W/OC; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 210813
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Injury (2348)
Event Date 01/18/2019
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
It was reported by the affiliate via email that during an arthroscopic bankart repair the surgeon implanted the first anchor and it backed out right after being implanted into the shoulder.Then, with the different bone hole created, the surgeon implanted the second anchor, which broke.Finally, the surgery was completed by using alternative anchor with another bone hole.There were no broken parts in the patient¿s body.It was brand new and the first use when the issue occurred.There was no harm to the patient.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.Hence, this complaint cannot be confirmed.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.Furthermore, a non-conformance search was conducted and no non-conformances were identified for this part number (210813), lot number (1l29572) combination.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRYPHON P BR DS ANCHOR W/OC
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8307878
MDR Text Key135143226
Report Number1221934-2019-56334
Device Sequence Number1
Product Code MAI
UDI-Device Identifier10886705001279
UDI-Public10886705001279
Combination Product (y/n)N
PMA/PMN Number
K150209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 01/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2021
Device Catalogue Number210813
Device Lot Number1L29572
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/21/2019
Initial Date FDA Received02/05/2019
Supplement Dates Manufacturer Received02/15/2019
Supplement Dates FDA Received02/21/2019
Patient Sequence Number1
-
-