Catalog Number 222330
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/18/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event Description
It was reported by the affiliate via email that during an arcr surgery the anchor backed out and got a crack on the tip.A second anchor was then implanted, which backed out.The surgery was completed using a third anchor with no patient harm.There were no broken pieces left in the patient.These anchors were brand new and first time use when this occurred.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint indicates that the device will not be returned and therefore not available for evaluation.No further information regarding the cause of the defect has been provided to help determine a root cause for this failure.This complaint cannot be confirmed.We cannot discern a root cause for the reported failure mode.No nonconformances were identified for this part number, serial number combination.Review conducted per qlik query executed on 2/13/2019.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.75 HEALIX ADVANCE KNTLSS BR
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8310357
MDR Text Key137147404
Report Number1221934-2019-56338
Device Sequence Number1
Product Code MAI
UDI-Device Identifier10886705023462
UDI-Public10886705023462
Combination Product (y/n)N
PMA/PMN Number
K130917
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 01/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Catalogue Number222330
Device Lot NumberL909348
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/20/2019
Initial Date FDA Received02/06/2019
Supplement Dates Manufacturer Received02/13/2019
Supplement Dates FDA Received02/15/2019
Patient Sequence Number1
-
-