SMITH & NEPHEW, INC. DBL OFFSET BROACH HDLE RT; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Catalog Number 71360090
Device Problem Misassembled (1398)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/16/2019
Event Type  malfunction  
Event Description
It was reported that broach handles separated from the broach when backing out from femur.No delay, backup available.
 
Manufacturer Narrative
The associated device was returned and evaluated.A visual / functional inspection of the returned device could not confirm the stated failure mode.The device functions as intended with a mating part.The device was manufactured in 2014 and exhibits signs of extensive wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Credit will be issued for the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DBL OFFSET BROACH HDLE RT
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8311443
MDR Text Key135202167
Report Number1020279-2019-00476
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010610362
UDI-Public03596010610362
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71360090
Initial Date Manufacturer Received 01/16/2019
Initial Date FDA Received02/06/2019
Supplement Dates Manufacturer Received01/16/2019
Supplement Dates FDA Received03/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-