• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE; EQUINOXE CAGE GLENOID SMALL, ALPHA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE; EQUINOXE CAGE GLENOID SMALL, ALPHA Back to Search Results
Catalog Number 314-13-02
Device Problem Disconnection (1171)
Patient Problem No Code Available (3191)
Event Date 01/24/2019
Event Type  Injury  
Manufacturer Narrative
Pending evaluation.
 
Event Description
Index surgery: (b)(6) 2019.Revision due to cage glenoid disassociating from the pegs without any traumatic event.
 
Manufacturer Narrative
Section h10: (b2) added check for "hospitalization - initial or prolonged" (d4) expiration date: 10-apr-2022, unique identifier (udi) #: (b)(4).(d5) operator of device: health_care_professional (g1) mdr reporting contact name: (b)(6).(g5) pma/510(k)number: k113309.(h3) the revision reported may have been the result of the glenoid component bottoming out on the glenoid wall upon implantation, possibly related to the small glenoid anatomy, which led to splaying of the posterior inferior peg and ultimate failure.However, the surgeon stated that the patient experienced failure of the rotator cuff which may have led to instability and, as a result, shear stress on the glenoid.Because the explanted device was not returned to exactech for evaluation and no additional x-ray images were provided, specifically an axillary view, it is not possible to determine the underlying cause for this case.(h4) device manufacture date: (b)(6) 2017.*no information provided in the following section(s): a1, a2, a3, a4, a5, b6, b7, e3, g8, h6, h7.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Type of Device
EQUINOXE CAGE GLENOID SMALL, ALPHA
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
MDR Report Key8311563
MDR Text Key135203540
Report Number1038671-2019-00065
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
PMA/PMN Number
K113309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/10/2022
Device Catalogue Number314-13-02
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/24/2019
Initial Date FDA Received02/06/2019
Supplement Dates Manufacturer Received10/30/2019
Supplement Dates FDA Received11/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-