It was reported that during the procedure, when the ventricular lead was connected to the device, ventricular pacing (vp) was not recorded.The status was not improved after attempting to put the ventricular lead into the device.Therefore, a check on psa was performed to see if this was caused by the lead or the device.When the v lead was connected to the pacemaker system analyzer (psa), the status was improved, and vp was recorded properly.After attempting to connect the v lead to the device again, the issue was not solved.The device was replaced with another device and no reported issue was observed when the v lead was connected to the 2nd device and vp was properly recorded without any problem.The procedure was completed without any adverse consequences to the patient.
|