Model Number 850000010 |
Device Problems
False Negative Result (1225); Component Missing (2306); Defective Device (2588)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/17/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This is our initial report on this incident.
|
|
Event Description
|
The customer reported false negative or weak antibody screening and identification results for several samples when compared with results generated with tube testing using the enhancement medium polyethylene glycol (peg) and the competitor instrument echo.In particular the customer reported a missed antibody when using biotestcell 3 lot #8850011-00 on two different tango infinity instruments.The customer stated that the specificity of the missed antibody was anti-jk(a).The customer did not return the patient sample that had caused a false negative test result but did sent in the complaint sample of biotestcell 3 and further reagents.Within its shelf life the retention sample of the supposedly defective lot of biotestcell 3 was tested by our quality control laboratory with different controls and samples, e.G.Anti-jk(a) on tango infinity.All positive and negative reactions were correct.We did not observe any false negative reaction.Testing of the reagents provided by customer is still ongoing.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.Regarding the affected tango infinity, the result images of sample in question were reviewed and the negative reactions could be confirmed also visually.The instrument was inspected by our field service engineer on january 23, 2019.He performed minor adjustments as well as coordinated teaching.Upon metrology, serology and titration testing, the instrument was confirmed to operate within specification.The metrology qualification procedure was completed on 01/23/2019 with passing results.Dilution studies were performed using the solidscreen ii qc positive control.The results were as expected.In some cases there was a "?" interpreted result for a strong positive reaction.This phenomenon is known and can be addressed by installing the sp4 upgrade.The customer then tested the sample on manual tube with both liss and peg.Tested with 2 jka homozygous cells from immuco pano-cell lot 48411, exp 2/1/2019 (cells 4 and 5).Cells 4 and 5 were negative in tube testing with liss.Cell 4 was negative and cell 5 was +/- in tube testing with peg.Both tango infinity instruments gave negative results.The log files did not show any issue relevant abnormalities.No indication for an instrument malfunction could be identified.The service engineer confirmed a proper function of the instrument by metrology qualification.As to the database, no problems with quality control testing on solidscreen ii method occurred.Since the samples were not available, we can only speculate that the antibodies were present in too low concentration and are borderline to be detected by the solidscreen ii method.Also, customer uses different methods, which are not comparable, such as the tube technique with peg enhancement.The instruction for use contains the following statement: negative reactions can occur in samples with excessively low antibody concentration because this test method cannot detect them.No test method can detect all blood group antibodies.
|
|
Event Description
|
The customer reported false negative or weak antibody screening and identification results for several samples when compared with results generated with tube testing using the enhancement medium polyethylen glycol (peg) and the competitor instrument echo.In particular the customer reported a missed antibody when using biotestcell 3 lot #8850011-00 on two different tango infinity instruments.The customer stated that the specificity of the missed antibody was anti-jk(a).During the course of the complaint handling the customer informed us about the false negative reaction of the anti-jk(a) with biotestcell-i11 plus (ref # 816022100, lot #8850011-00, expiry date 2019-02-04.At the time the customer informed us, the supposedly defective lot of biotestcell-i11 plus was already expired.Therefore our quality control laboratory did not test their retention sample.The customer did not return the patient sample that had caused a false negative test result but the complaint sample, two different lots of anti-human globulin (ahg) anti-igg solidscreen ii and mlb2.At the time our quality control laboratory received the reagents provided by the customer, the supposedly defective lot of biotestcell 3 was already expired.But the retention sample of the allegedly defective lot of biotestcell 3 was tested in our qc laboratory within its shelf life with different controls and samples, e.G.An anti-jk(a) on tango infinity.All positive and negative reactions were correct.We did not observe any false negative reaction.Furthermore both ahg reagents provided by the customer were tested for potency and specificity and met the acceptance criteria.The mlb2 was also tested and met the acceptance criteria.Testing by our quality control laboratory confirmed that the allegedly defective lot of biotestcell 3 functions correctly.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.Regarding the affected tango infinity, the result images of sample in question were reviewed and the negative reactions could be confirmed also visually.The instrument was inspected by our field service engineer on january 23, 2019.He performed minor adjustments as well as coordinated teaching.Upon metrology, serology and titration testing, the instrument was confirmed to operate within specification.The metrology qualification procedure was completed on 01/23/2019 with passing results.Dilution studies were performed using the solidscreen ii qc positive control.The results were as expected.In some cases there was a "?" interpreted result for a strong positive reaction.This phenomenon is known and can be addressed by installing the sp4 upgrade.The customer then tested the sample on manual tube with both liss and peg.Tested with 2 jka homozygous cells from immuco pano-cell lot 48411 exp 2/1/19 (cells 4 and 5).Cells 4 and 5 were negative in tube testing with liss.Cell 4 was negative and cell 5 was +/- in tube testing with peg.Both tango infinity instruments gave negative results.The log files did not show any issue relevant abnormalities.No indication for an instrument malfunction could be identified.The service engineer confirmed a proper function of the instrument by metrology qualification.As to the database, no problems with quality control testing on solidscreen ii method occurred.Since the samples were not available, we can only speculate that the antibodies were present in too low concentration and are boarderline to be detected by the solidscreen ii method.Also, customer uses different methods, which are not comparable, such as the tube technique with peg enhancement.The instruction for use contains the following statement: negative reactions can occur in samples with excessively low antibody concentration because this test method cannot detect them.No test method can detect all blood group antibodies.
|
|
Manufacturer Narrative
|
This is our final report on this incident.
|
|
Search Alerts/Recalls
|
|