HOLOGIC, INC SELENIA/LORAD MAMMOGRAM UNIT; FULL FIELD DIGITAL SYSTEM, XRAY, MAMMOGRAPHIC
|
Back to Search Results |
|
Device Problems
Device Displays Incorrect Message (2591); Connection Problem (2900)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/06/2019 |
Event Type
malfunction
|
Event Description
|
Patient underwent mammogram and machine stopped functioning during exam.An exposure and communication error was noted.Upon investigation, a cable was replaced and a loose connection was found.No patient harm.
|
|