• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: (b)(4) 2019.International udi: (b)(4).Reporter phone number unknown.The philips service engineer (se) inspected the device.The se replaced the central processing unit (cpu) printed circuit board assembly (pcba) and the cpu tray cover.The ventilator passed all testing.
 
Event Description
It was reported that the ventilator generated a primary alarm failed error code.There was no patient involvement.The event date was not specified; estimate used.
 
Manufacturer Narrative
Date rec'd by mfr: 26mar2019.Information was received that the cpu (central processing unit) tray cover was replaced due to it being damaged.A cpu pcba (printed circuit board assembly) was returned for analysis.A visual inspection of the returned component was performed, no notable conditions were found.The returned component was installed into a test ventilator for analysis.An investigation was performed and the product analysis technician reported that no fault was found.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8348087
MDR Text Key136606666
Report Number2031642-2019-00969
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/27/2019
Initial Date FDA Received02/18/2019
Supplement Dates Manufacturer Received01/27/2019
Supplement Dates FDA Received04/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-