The hyperform occlusion balloon catheter was returned for analysis.The hyperform guidewire was not returned as it was discarded.Upon visual inspection, no issues or irregularities were found with the hyperform hub.No bends or kinks were found with the hyperform catheter body.An attempt was made to flush the catheter lumen with water; however, was found to be occluded with what is likely to be dried contrast.In order to test for balloon inflation, the balloon subassembly was separated (cut) from the catheter body.A mandrel was inserted into the distal tip of the balloon for leak test.An attempt was made to inflate the balloon; however, the balloon was found to be leaking.Upon microscopic examination, the balloon was found to be ruptured at the leak.Based on the device analysis and reported information, the report of ¿no/slow inflation during set up¿ was confirmed.The appearance of the balloon is consistent with failure due to over-inflation.Rupture can occur when the balloon is inflated beyond the rated volume or due to inflation rate.However, the cause for the balloon rupture could not be conclusively determined.A device history record review did not identify any anomalies associated with this event.These types of events will be monitored as part of post market vigilance complaint trending and monitoring.Per instructions for use (ifu), ¿to inflate the balloon, use the 1 cc syringe to gently infuse the desired volume of recommended contrast solution.Balloon cannot inflate unless a portion of the distal 10 cm of the guidewire tip is occluding the distal inflation holes of the catheter.Do not exceed the maximum recommended inflation volume as balloon rupture may occur".If information is provided in the future, a supplemental report will be issued.
|