• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: (b)(4) 2019.International udi: (b)(4).The philips field service engineer replaced the oxygen solenoid valve to resolve this issue.The unit passed the performance verification tests when the repair was complete.
 
Event Description
Philips field service engineer (fse) reports o2 concentration failure.There was no patient or user harm reported.The event date was not specified; estimate used.
 
Manufacturer Narrative
Date rec'd by mfr: 10may2019.Date of report: 12jun2019.The oxygen solenoid was received for evaluation.Excessive lubricant on the solenoid body was noted during visual inspection.After testing, it was determined that the reported failure was due to the leaking solenoid that had debris between sealant o-ring and body.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8350294
MDR Text Key136733827
Report Number2031642-2019-00978
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 01/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/28/2019
Initial Date FDA Received02/19/2019
Supplement Dates Manufacturer Received01/28/2019
Supplement Dates FDA Received06/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-