• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 01/25/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova deutschland manufactures the heater-cooler system 3t.The event occurred in (b)(6).A review of the dhr could not identify any deviations or nonconformities relevant to the issue.A picture of a large metallic particle has been provided and investigation is ongoing.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event Description
Livanova deutschland received a report of a heater-cooler system 3t with large metallic particles found in the drain bucket during water change.There was no report of patient injury.
 
Manufacturer Narrative
H.10: through follow-up communication with the customer, livanova deutschland learned that the particles were identified during the user maintenance only.The facility returns the device in use only once they are completely removed from the device by flushing the tubes.The reported event has been re-evaluated as non reportable.
 
Event Description
See initial.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8350705
MDR Text Key138813358
Report Number9611109-2019-00093
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/25/2019
Initial Date FDA Received02/19/2019
Supplement Dates Manufacturer Received04/25/2019
Supplement Dates FDA Received04/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-