• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 02/19/2019.(b)(4).The philip's product support engineer (pse) replaced the flow sensor assembly to resolve the reported issue.The pse also found the screw supports on the front and rear bezels were broken.The pse replaced the front and rear bezels.The ventilator passed all testing and operated within the manufacturing specifications.
 
Event Description
The customer contacted philips technical support (ts) stating that the unit had a flow sensor failure during performance verification testing (pvt).The customer reported there was no patient involvement.The event date was not specified; estimate used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RICA
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
lisa cardenas
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key8350961
MDR Text Key136616401
Report Number2031642-2019-00994
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/28/2019
Initial Date FDA Received02/19/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-