According to the complaint received, observed deviation on the na+ and k+ parameter on an abl90 flex analyzer compared to the measurements from the central laboratory.The measurements are shown below, demonstrating a negative deviation of 9 mmol/l for na+ and negative deviation of 0.8 mmol/l for k+.On (b)(6) 2019 12:21 abl90 flex: na+ 128 mmol/l; (b)(6) 2019 12:21 abl90 flex: k+ 3.8 mmol/l; central laboratory: na+ 137 mmol/l; central laboratory: k+ 4.6 mmol/l.Based on the measurements listed, the customer reported the na+ and k+ result from the abl90 flex analyzer as false low.
|