• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE TMZF HIP STEM #5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE TMZF HIP STEM #5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6020-0537
Device Problems Degraded (1153); Material Deformation (2976)
Patient Problem Injury (2348)
Event Date 01/24/2019
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event Description
Stryker rep reported that they were in a case where the surgeon was revising an accolade with a 44mm lfit head.The trunnion was badly worn and metal debris was present.
 
Manufacturer Narrative
An event regarding trunnionosis involving an unknown accolade stem was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review: review indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: review indicated there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information, including device detail, operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event Description
Stryker rep reported that they were in a case where the surgeon was revising an accolade with a 44mm lfit head.The trunnion was badly worn and metal debris was present.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE TMZF HIP STEM #5
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8358933
MDR Text Key136797411
Report Number0002249697-2019-01134
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
PMA/PMN Number
K121308
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 05/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2011
Device Catalogue Number6020-0537
Device Lot Number18498803
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/24/2019
Initial Date FDA Received02/21/2019
Supplement Dates Manufacturer Received04/11/2019
Supplement Dates FDA Received05/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-