Unspecified date: patient presented to the facility in an ambulance after a high speed mva roll over.The patient required trama series x-ray's and ct's.The patient was tested 2x on the hcg 10/20 pregnancy test device and both results were negative.Ct was performed based on the false negative results.Confirmatory blood work was also performed, and the result was 110 miu/ml.Result of confirmatory blood work was not obtained until after the ct was performed.The patient was admitted to the facility for observation overnight and discharged the next day.Customer states the patient is still considered to be pregnant.No adverse outcomes reported.Although further information was requested, no further information was provided.
|
Investigation conclusion: an investigation was performed on retention product from the reported lot number.Retention devices were tested with qc cutoff standard (20 miu/ml) and middle positive standard (100 miu/ml).Results were read at 3 minutes and all devices yielded expected positive results.No false negative results were observed during in-house testing.Manufacturing batch record review did not uncover any abnormalities and found that the lot met quality control specifications.The complaint history for this lot was reviewed and no indications of a systemic issue were identified.A root cause could not be determined from the available information, as retention product performed as expected during in-house testing.Please note, very dilute urine specimens, as indicated by a low specific gravity, may not contain representative levels of hcg.If pregnancy is still suspected, a first morning urine specimen should be collected 48 hours later and tested.This test provides only a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.
|