(b)(4).Batch # r93780.Investigation summary: the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched and had evidence of contact with metal in or out of the operative field.This blade tip portion may have broken off of the device during transport to our analysis site.During functional testing on gen11 an alert screen was displayed.A probable cause for the device to stop activating and for the gen11 to display an alert screen is blade damage.The device was analyzed and it was determined that it was possibly used in more than one procedure, based on generator data.The device is intended and labeled for single patient use.If the device is connected to multiple generators an alert screen will be displayed indicating ¿adaptive tissue technology features are not available in this device¿.The device was disassembled to inspect the internal components and no anomalies were noted.Due to the multiple generator usage, which is evidence of possible reuse, we are unable to determine how this condition impacted the performance of the device and therefore cannot conclude root cause.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, contact with staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activation may increase the severity of the blade damage.This, in turn, can result in the device failing the pre-run test with the generator and displaying an alert screen.The alert screens that can result may include ¿tighten assembly,¿ ¿blade error detected,¿ or "relax pressure on blade," followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damaged blade can result in a broken blade.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
It was reported prior to a laparoscopic cholecystectomy procedure, after connecting and doing testing, a tighten assembly error occurred.The device was not used on patient, error occurred in beginning.There was no patient consequence.
|