• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET FRANCE S.A.R.L. OPTIPAC 40 REFOBACIN BONE CEMENT R-3; BONE CEMENT, ANTIBIOTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET FRANCE S.A.R.L. OPTIPAC 40 REFOBACIN BONE CEMENT R-3; BONE CEMENT, ANTIBIOTIC Back to Search Results
Catalog Number 4710500394-3
Device Problem Improper Chemical Reaction (2952)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/29/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation, [product was discarded].The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It has been reported that during surgery the cement harden time of the optipac refobacin bone cement varies from just a few minutes to 20 minutes, although the products are stored to room temperature and placed one next to each other.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following section has been updated : d2, g1-2, g4, h1/h2/h3, h6, h10.The product analysis can't be performed as the product was not returned.The documentary review showed that products were manufactured according to the pre-defined specifications of biomet france.A retain sample of batch involved in this complaint has been tested in the laboratory under standardized conditions (23°c; rel.Humid >=40 %).No unusual behavior during mixing, handling or setting.The reported behaviour of the cement cannot be reproduced.As explained in the ifu, the working time and the rate of polymerization are strongly dependent on ambient temperature and temperature of the components.The hardening time is shorter at high temperatures and longer at low temperatures.According to available data, the exact root cause can¿t be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It has been reported that during surgery the cement harden time of the optipac refobacin bone cement varies from just a few minutes to 20 minutes, although the products are stored to room temperature and placed one next to each other.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIPAC 40 REFOBACIN BONE CEMENT R-3
Type of Device
BONE CEMENT, ANTIBIOTIC
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
MDR Report Key8369293
MDR Text Key137138878
Report Number3006946279-2019-00149
Device Sequence Number1
Product Code MBB
UDI-Device Identifier04040029922347
UDI-Public(01)04040029922347
Combination Product (y/n)N
PMA/PMN Number
K171540
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Catalogue Number4710500394-3
Device Lot Number811AA09170
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/30/2019
Initial Date FDA Received02/26/2019
Supplement Dates Manufacturer Received07/18/2019
Supplement Dates FDA Received07/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-