(b)(4).This follow-up report is being submitted to relay additional information.The following section has been updated : d2, g1-2, g4, h1/h2/h3, h6, h10.The product analysis can't be performed as the product was not returned.The documentary review showed that products were manufactured according to the pre-defined specifications of biomet france.A retain sample of batch involved in this complaint has been tested in the laboratory under standardized conditions (23°c; rel.Humid >=40 %).No unusual behavior during mixing, handling or setting.The reported behaviour of the cement cannot be reproduced.As explained in the ifu, the working time and the rate of polymerization are strongly dependent on ambient temperature and temperature of the components.The hardening time is shorter at high temperatures and longer at low temperatures.According to available data, the exact root cause can¿t be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|