• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC. FLUENT FLUID MANAGEMENT SYSTEM; HYSTEROSCOPIC INSUFFLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC. FLUENT FLUID MANAGEMENT SYSTEM; HYSTEROSCOPIC INSUFFLATOR Back to Search Results
Model Number FLT-112S
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/29/2019
Event Type  malfunction  
Manufacturer Narrative
Lot and serial number of the disposable device not provided by the complainant, therefore the expiration date is not known.The device is not being returned therefore, a failure analysis of the complaint device cannot be completed.Lot number of the disposable device not provided by the complainant, therefore the manufacture date is not known.Device history record (dhr) review and sterile lot review were unable to be conducted for the disposable device as the lot number was not provided by the complainant.Internal complaint reference: (b)(4).The fluent system was returned for evaluation and passed all functional testing.
 
Event Description
It was reported that when using the fluent system a reprime had to be completed because the waste bag hooks didn't sense the bag, and the system was alarming with the screen flickering between "reprime" and "operating".During cutting it was noted that water was dripping from the tissue canister and the specimen was coming out of the sock and plugging up the mesh.Due to the weight and pressure the canister broke off and the fluid and specimen spilled to the floor.The canister was not able to be used since the specimen had plugged the mesh holes and a basin was used to catch the outflow and the specimen was then strained out.The fluent system was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENT FLUID MANAGEMENT SYSTEM
Type of Device
HYSTEROSCOPIC INSUFFLATOR
Manufacturer (Section D)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
kristin fornieri
36 & 37 apple ridge road
danbury, CT 06810
2037318491
MDR Report Key8370311
MDR Text Key137417597
Report Number1222780-2019-00039
Device Sequence Number1
Product Code HIG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K180825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFLT-112S
Device Catalogue NumberFLT-112S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/29/2019
Initial Date FDA Received02/26/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-