Catalog Number L96409
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Not Applicable (3189)
Event Date 02/05/2019
Event Type  Injury  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.
 
Event Description
While doing an endo hip with corail cemented stem,the surgeon tried to reduce hip with trials and the femur broke.Nothing major, just a small crack proximally, the surgeon used a cable to fix fracture surgery was extended about 15 minutes in order to put cable in.No harm to patient other than a cable installed.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.No device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot = null.Device history batch = null.Device history review = null.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL CEM STEM STD S9
Type of Device
CORAIL AMT CEMENTED IMPLANTS : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS 3003895575
7 allée irène joliot curie
bp 256
saint priest cedex 69801
FR  69801
MDR Report Key8374852
MDR Text Key137292550
Report Number1818910-2019-85926
Device Sequence Number1
Product Code JDG
Combination Product (y/n)N
PMA/PMN Number
UNK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberL96409
Device Lot NumberD18092861
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/05/2019
Initial Date FDA Received02/27/2019
Supplement Dates Manufacturer Received03/07/2019
Supplement Dates FDA Received03/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-