• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG SEMI-RIGID 6º URETEROSCOPE W/STRAIGHT; SEMI-RIGID SCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG SEMI-RIGID 6º URETEROSCOPE W/STRAIGHT; SEMI-RIGID SCOPE Back to Search Results
Model Number 27011L
Device Problem Material Frayed (1262)
Patient Problems Excessive Tear Production (2235); Patient Problem/Medical Problem (2688)
Event Date 01/17/2019
Event Type  malfunction  
Manufacturer Narrative
We evaluated the subject semi-rigid ureteroscope and we found the distal end is a little rough from long uses.Scopes shaft is bent and with approximately 60% broken light fibers.The seni-rigid ureteroscope was in use for approximately 12 years, purchased (b)(6) 2007.
 
Event Description
Ureteroscope was placed into ureter and advanced toward the kidney.During removal, the ureter stuck to scope and was avulsed.The ureter could not be repaired and a nephrectomy was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEMI-RIGID 6º URETEROSCOPE W/STRAIGHT
Type of Device
SEMI-RIGID SCOPE
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen,
GM 
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen,
GM  
Manufacturer Contact
susie chen
2151 east grand avenue
el segundo, CA 90245-5017
MDR Report Key8379098
MDR Text Key138408957
Report Number9610617-2019-00021
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04048551231371
UDI-Public4048551231371
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K940464
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27011L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/28/2019
Initial Date FDA Received02/28/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/03/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age62 YR
Patient Weight127
-
-