Siemens filed the initial mdr 1219913-2019-00031 on (b)(6) 2019.(b)(6) 2019 additional information: the cse replaced the wash separation manifold and tested the system.A possible cause of the elevated results for this direct assay are poor aspiration of waste or system contamination.For this assay the wash separation manifold is responsible for the aspiration of unbound waste material at aspirate positions 1, 2, 3, and 4 and dispensing of wash 1 at dispense ports 1, wash displacement (wd), and 3 during the washing process.Damage to the wash separation manifold could cause elevated results.However, this could not be confirmed as the cause of the false positive results.The instrument is performing within specifications.No further evaluation of the device is required.Mdr 1219913-2019-00027 supplemental report 1 was filed for the same event (different patient).
|