Model Number 10-80-00 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/05/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was not provided.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 roller pump displayed an incorrect flow during procedure.There was no report of patient injury.
|
|
Event Description
|
See initial.
|
|
Manufacturer Narrative
|
A livanova field service representative was dispatched onsite to investigate and could confirm the reported issue.He found out that the switches on the motor control board were not in the correct position.This led to the reported event.The technician set the switches in the correct position and the issue was solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
|
|